Single Monitor
Double Monitor Digital 25
For Live and stored – image fluoroscopy before, during and after surgery, the BV25 has the capability of storing images for video hard copy documentation. Also provide radiographic documentation when required.
X-ray generator: DC converter – generator.
X-ray tubes – focal spots: dual-focus stationary anode 0.6 and 1.8
Image intensifier: 15cm (6 in)high quantum absorption model with fibre output.
TV camera tube: VIDICON with fibre input.
Optical coupling between image intensifier and TV camera tube: glass-fibre optics.
Source of image distance: 90 cm.
Dose rate control: Automatic, simultaneous kV/mA control with lock and manual override facilities.
KV/mA range: from 0.1 mA at 40 kV to 3 mA at 100 kV.
Focal spot:06.
Grid: Fixed circular grid 44 lines/cm, ratio 8, SID= 90 cm.
Iris Diaphragm: electronically controlled iris diaphragm, automatically limited to size of image intensifier input field (15 cm 0).
Remotely controlled and steplessely adjustable to a field size of 5 cmdiameter at input field image intensifier.
Shutters: two parallel shutters of 0.5 mm Cu, remotely controlled and steplessely adjustable to a field of 1 to 16 cm wide at range intensifier.
The shutters can be rotated+ 90 degree, in the mid position, the slit aperture is perpendicular to the length of the generator housing.
XTV 5 50 Hz: 625 lines,
60 Hz: 525 lines.
Bandwidth:15 MHz,
Control: AGC feedback to dose rate control system.
Video monitor(s): 51 cm (20”)screen motorized, remotely controlled image rotation and revesal. Output signal for recording system: outpout level of the composite signal always 1.o Vp-p positive across 75 ohms.
Measuring field selection for adapting control mode to the objects to be examined: small measuring field: 0.3 X diameter of TV image, e.g. extremities.
Intermediate measuring field: 0.5 x diameter of TV image, e.g. bile duct and gall bladder. The measuring field is selected automatically as a function of the k V used.
Height adjustment: 50 cm, motorized.
Horizontal Movement: 20 cm.
Scanning in horizontal plane:12*1/2 degree to either side.
Rotation of c-arm: 205 degree in either direction.
Orbital movement of c-arm: 115 degree,
Distance between x-ray tube space and image intensifier with 30 cm spacer is: 60 cm.
Tube current: fixed. 20 mA .
Tube voltage: adjustable in steps of 5 kV from 40 to 100 kV.
Exposure time: adjustable in steps of 10 ms to 4.0 s(mAs is displayed).
Focus: 1.8.
Radiogarphy parameter display: digital display of tube voltage (kV) and exposure time/current product (mAs).
Collinator: If iris diaphragm is set to be less than 15 cm during fluoroscopy, this remains so unless cancelled. Circular fields of 15 or 24 cm can be selected by remote control.
Rotatable cassette holder:suitable for cassette and grid cassettes of 24 x 24 cm.
Proffered grid:ratio 10.44 lines/cm, SID = 90 cm.
Inherent filtration:3mm A1 at 75 kV.
Leakage radiation factor:100 kVdc,0.06 mA (2160 mAs/hr)
Fluoroscopy:voltage deviation + 6% + 2kV , current deviation +7% +0.02 mA.
Radiography: voltage deviation + 10%, current deviation +10%, time deviation +2ms +3%, mAs product deviation + 12% +0.04 mAs.
C-arm stand: 210 kg. Monitor trolley exculiding image memory single monitor:131 kg,
dual monitors: 164 kg.