Height: 32.6 cm (12.8”). Width: 32 cm (12.5”). Depth: 14.1 cm (5.6”). Weight: 17.9 lbs
Size: 10.4” diagonal.
Type: monochrome high-bright electroluminescent and color thin
film transistor liquid crystal display.
Resolution: 640 x 480 pixels.
Number of traces: 6.
Length of trace: 6.0 at 25 mm/sec.
Sweep speed: 25 mm/s, all waveforms except CO2.
Waveform display options: full or individual.
Information window: displays all non-realtime
information without obstructing the display of real-time information.
Display
organization: prioritized by parameter.
Classification:crisis, warning, advisory, and message.
Notification: audible and visual.
Setting: default and individual.
Silencing: 1 minute, current alarm only.
Volume: default
70%, 70 dB measured at 1 meter.
Leads: I, II, III, V, aVR, aVL, and aVF.
Leads analyzed simultaneously: I, II, III and V
(multi-lead mode).
Heart rate range: 30 to 300 bpm.
Measurement technique: impedance variation detection.
Respiration rate range: 1-200
breaths per minute.
Alarms: user-selectable upper and lower limits, and user-selectable
apnea limit.
Parameters monitored: arterial oxygen saturation (SpO2) and peripheral pulse rate (PPR).
Range: 50-100%;
Accuracy: ±2% (70-100%), ±3% (50-60%).
PPR – range: 20-
250 bpm;
Accuracy: ±3 bpm.
Alarms: user-selectable upper and lower limits for SpO2
and PPR.
Channels: 2.
Range: 0°C to 45°C (32°F to 113°F.
Resolution: ±0.1°C.
Parameters
displayed: T1 and T2.
Alarms: selectable upper and lower limits for T1 and T2.
Measurement Technique: oscillometric.
Displayed parameters: systolic, diastolic, and
mean pressures, pulse rate, time of last measurement.
Measurement modes: manual, auto
and stat.
Heart rate detection: 30 to 300 beats per minute.
Alarms: user-selectable upper
and lower limits for systolic, diastolic, and mean pressures.
Power Requirements 90-132 VAC 50/60 Hz; 190-264 VAC 50/60 Hz.
Power consumption: 90-132 VAC 2A; 190-264 VAC 50/60 1A